The AMETEK RTCt-157 is a high-precision reference temperature calibrator for professional temperature calibration in the range of −45 °C to 155 °C at 23 °C ambient temperature. This powerful dry-block calibrator combines exceptionally high temperature stability with modern touchscreen operation, automated calibration functions and extensive options for integrating external reference probes.
With a specified temperature stability of just ±0.005 °C, the AMETEK RTCt-157 is particularly suitable for demanding applications in calibration laboratories, quality assurance, the pharmaceutical and food industries, research and industrial temperature measurement. With an external JOFRA STS-200 reference probe, the system achieves an accuracy of up to ±0.040 °C relative to the reference. The specified total accuracy, including STS sensor and accredited system calibration, is ±0.065 °C.
The modern RTCt generation extends the proven AMETEK JOFRA RTC platform with touchscreen operation, wireless communication, real-time graphics and – with the corresponding model – the ability to calibrate two devices under test simultaneously.
europascal supports you in selecting the right model as well as suitable reference probes, DLC sensors, inserts and accessories for your individual temperature calibration.
Your Benefits at a Glance
- Temperature range from −45 °C to 155 °C at 23 °C ambient temperature
- Extended range down to −57 °C at 0 °C ambient temperature
- Very high temperature stability of ±0.005 °C
- Accuracy up to ±0.040 °C with external STS-200 reference probe
- Total accuracy up to ±0.065 °C including reference probe and accredited system calibration
- Internal reference accuracy of ±0.11 °C
- Large immersion depth of 160 mm
- Active dual-zone temperature control
- Patented DLC Dynamic Load Compensation
- Modern touchscreen user interface
- Real-time graphics for monitoring stability and temperature trends
- Wireless communication and browser-based access
- Two devices under test measurable simultaneously with model B
- Auto Step function with up to 20 temperature steps
- Set Follows True function with models B and C
- Automatic temperature switch testing
- Intelligent JOFRA reference probes with stored calibration data
- Display resolution down to 0.001 °C
- Display in °C, °F or K
- JofraCal calibration software
- Suitable for laboratory and mobile use
AMETEK offers the RTCt-157 in versions A, B and C. Model B has inputs for a reference sensor, a DLC sensor and two devices under test. Model C has inputs for a reference sensor and a DLC sensor.
Ideal for Demanding Calibration Applications
The AMETEK RTCt-157 was developed for users with very high requirements for accuracy, stability and reproducibility. The temperature range of −45 °C to 155 °C covers numerous applications that require temperatures both well below room temperature and elevated temperatures.
Typical fields of application include calibration laboratories, in-house calibration laboratories, quality assurance, pharmaceuticals, biotechnology, the food and beverage industry, aerospace, automotive, power generation, mechanical engineering, production as well as research and development.
AMETEK explicitly names pharma, food and beverage, aerospace, automotive and power generation as typical applications of the RTCt-156 and RTCt-157 series.
The wide temperature range allows, for example, the calibration of Pt100 resistance thermometers, Pt1000 sensors, thermocouples, process probes, temperature transmitters and complete temperature measurement chains.
Flexible Calibration
The AMETEK RTCt-157 works as a high-precision dry-block calibrator. The device under test is placed in a suitable metal insert, which the powerful control system brings to the set temperature.
The immersion depth of 160 mm provides good conditions for sufficient thermal coupling of a wide variety of temperature probes.
Various inserts are available for different probe diameters. Multi-hole inserts also allow several temperature probes to be inserted at the same time.
The RTCt-157 B version extends these possibilities even further: two sensors under test can be connected electrically to the calibrator at the same time. This allows two temperature probes to be calibrated in parallel during a single temperature cycle, reducing the time required, especially for recurring calibrations.
Technical Highlights
At 23 °C ambient temperature, the AMETEK RTCt-157 covers a range of −45 °C to 155 °C. At an ambient temperature of 0 °C, the lower limit extends to −57 °C. At 40 °C ambient temperature, the specified range is −31 °C to 155 °C.
An outstanding feature is the temperature stability of ±0.005 °C. AMETEK specifies this value after an additional stabilization time of 15 minutes and over a measuring period of 30 minutes.
With the internal reference, the accuracy is ±0.11 °C. With versions B and C, an external JOFRA STS-200 reference temperature probe can be used, giving an accuracy relative to the reference of ±0.040 °C. For the complete system including STS reference probe and accredited system calibration, ±0.065 °C is specified.
Technical Data
| Feature | Value |
|---|---|
| Manufacturer | AMETEK JOFRA |
| Model | RTCt-157 |
| Device type | Reference temperature calibrator |
| Calibration principle | Dry block |
| Temperature range at 23 °C | −45 °C to 155 °C |
| Temperature range at 0 °C | −57 °C to 155 °C |
| Temperature range at 40 °C | −31 °C to 155 °C |
| Stability | ±0.005 °C |
| Accuracy, internal reference | ±0.11 °C |
| Accuracy with external STS-200 | ±0.040 °C |
| Total accuracy with STS-200 | ±0.065 °C |
| Radial uniformity | 0.01 °C at −32 °C |
| Display resolution | 1 °C, 0.1 °C, 0.01 °C or 0.001 °C |
| Temperature units | °C, °F and K |
| Immersion depth | 160 mm |
| Heating time −45 °C to 23 °C | approx. 7 minutes |
| Heating time 23 °C to 155 °C | approx. 17 minutes |
| Cooling time 155 °C to 100 °C | approx. 5 minutes |
| Cooling time 100 °C to 23 °C | approx. 10 minutes |
| Cooling time 23 °C to −30 °C | approx. 17 minutes |
| Cooling time −30 °C to −45 °C | approx. 20 minutes |
| Total cooling time 155 °C to −45 °C | approx. 52 minutes |
| Typical time to stability | approx. 10 minutes |
| Versions | RTCt-157 A, RTCt-157 B, RTCt-157 C |
| Dimensions L × W × H | 362 × 171 × 363 mm |
| Weight | approx. 10.3 kg |
Precise Temperature Calibration
The AMETEK RTCt-157 is designed in particular for applications that require not just a stable temperature source but a high-quality reference for professional comparison calibration.
For a calibration, the device under test is placed in a suitable insert and the desired set point is entered. Once the temperature and the defined stability conditions have been reached, the reference value and the reading of the device under test can be compared.
With a resolution down to 0.001 °C, even small changes in block temperature can be displayed clearly. This high resolution should not be confused with the measurement uncertainty or accuracy of the overall system, but it makes it easier to assess the stability behavior.
Active Dual-Zone Temperature Control
The AMETEK RTCt-157 features active dual-zone control, in which different areas of the calibration block are controlled independently of each other.
The aim is the most homogeneous temperature zone possible within the insert. Controlled axial temperature distribution can be decisive, especially with different probe lengths, several devices under test or sensors with high heat dissipation.
AMETEK combines this control with Dynamic Load Compensation, so that the effects of insert loading can be better monitored and compensated.
DLC Dynamic Load Compensation
DLC stands for Dynamic Load Compensation. This AMETEK technology takes into account thermal influences that can result from the number, size and design of the inserted temperature probes.
An additional DLC sensor monitors the temperature distribution within the insert. Together with dual-zone control, the system can react to changes in thermal load.
This is advantageous, for example, when several temperature probes are calibrated at the same time or when sensors with a relatively large diameter are calibrated.
AMETEK highlights the combination of dual-zone control and DLC especially for large sensors and multiple calibrations.
Set Follows True
Versions RTCt-157 B and RTCt-157 C support the Set Follows True function.
Here, the temperature of an external reference probe is used directly for control. The calibrator’s set point is adjusted automatically so that the temperature actually measured by the reference probe matches the desired calibration point as closely as possible.
This allows the reference temperature directly next to the device under test to be used for control.
This is particularly useful for demanding multi-point calibrations and applications with small permissible measurement uncertainties.
Intelligent Reference Temperature Probes
The AMETEK RTCt-157 supports intelligent JOFRA reference probes.
Their intelligent connectors store, among other things, the calibration data, coefficients, serial number and calibration date of the reference probe. This creates a largely automated plug-and-play system.
This reduces the effort required to set up a reference probe and at the same time lowers the risk of accidentally using incorrect sensor coefficients.
Automatic Calibration Routines
With Auto Step, the AMETEK RTCt-157 supports up to 20 programmable temperature steps.
Dwell times can be defined for each calibration point. The calibrator then approaches the programmed temperature values one after the other, so multi-point calibrations can be carried out with reduced operator effort.
This function is particularly suitable for recurring tests with identical calibration points.
Together with the JofraCal software, calibration processes can be further automated and documented.
Touchscreen and Real-Time Graphics
The RTCt generation features a modern touchscreen user interface. Measured values, set temperature, reference temperature and stability information can be displayed directly on the instrument.
A special feature is the real-time graph: as soon as the stability criteria are met, the display automatically adjusts its resolution, so very small temperature fluctuations are shown enlarged.
This visualization makes it easier to judge whether the calibrator is stable enough for taking readings.
Wireless Communication
The AMETEK RTCt-157 supports wireless access via a browser interface.
Calibration processes can thus be monitored and analyzed remotely, so the user does not have to stay right next to the temperature calibrator.
This is an advantage, for example, in calibration laboratories with several calibration stations running at the same time or during longer automatic test sequences.
Versions of the AMETEK RTCt-157
The RTCt-157 is available in three versions.
RTCt-157 A is the reference temperature calibrator without additional sensor inputs.
RTCt-157 B has inputs for an external reference sensor, a DLC sensor and two devices under test, so two sensors can be measured electrically and calibrated at the same time.
RTCt-157 C has inputs for an external reference sensor and a DLC sensor.
The equipment can thus be selected to suit the respective application and the existing laboratory infrastructure.
Calibration Service from europascal
In addition to professional temperature calibration equipment, europascal supports you in ensuring reliable and traceable measurement results in the long term.
Through the europascal Calibration Service, temperature probes, temperature measuring instruments and calibration equipment can be checked and calibrated regularly.
Regular calibration supports the metrological traceability of your test equipment and allows changes and drift to be detected at an early stage.
europascal also supports you in selecting suitable reference temperature calibrators, reference probes, DLC sensors and inserts, and with questions about application, maintenance and calibration.
Why Choose the AMETEK RTCt-157?
The AMETEK RTCt-157 combines an exceptionally wide low-temperature range with very high stability and modern automation.
The minimum temperature of −45 °C at 23 °C ambient temperature makes the RTCt-157 particularly attractive for applications where conventional dry-block calibrators with lower cooling capacity are not sufficient.
The stability of ±0.005 °C and accuracy of up to ±0.040 °C with an external STS-200 reference probe provide excellent conditions for demanding comparison calibrations.
Touchscreen, real-time graphics, wireless access, Auto Step, Set Follows True and DLC complement the metrological performance with functions that are especially beneficial in modern calibration laboratories.
Typical Calibration Applications with the AMETEK RTCt-157
The AMETEK RTCt-157 is particularly suitable for calibrating Pt100 resistance thermometers, Pt1000 sensors, thermocouples, industrial temperature probes, process probes, temperature transmitters and complete temperature measurement chains.
Depending on the version, multi-point calibrations, automated temperature tests and the simultaneous calibration of several sensors are also possible.
The wide range down to −45 °C makes the RTCt-157 particularly attractive for applications in the pharmaceutical and food industries, HVAC and refrigeration, automotive, aerospace, power generation, research as well as accredited and industrial calibration laboratories.
Thanks to the combination of a wide temperature range, high stability, external reference measurement, active dual-zone control and modern operation, the AMETEK RTCt-157 is a powerful solution for professional temperature calibration.
Manufacturer information on the AMETEK RTCt-157 can be found on the official AMETEK website.