europascal

Additel ADT 783-1K Precise Pressure Controller

The Additel ADT 783-1K Pressure Controller is a high-performance automated pneumatic pressure controller for professional calibration laboratories and test systems, covering pressures from approximately -0.99 to 70 bar (-14.35 to 1,000 psi). Depending on the selected ADT151 pressure module, accuracy is available down to 0.01 % FS or 0.01 % of reading, while control stability is <0.003 % FS with typical performance of 0.001 % FS. Two interchangeable internal pressure modules, fast pressure control in less than 10 seconds, automatic range switching and optional gauge-to-absolute pressure capability provide outstanding flexibility for automated pressure calibration.

The Additel ADT 783-1K Pressure Controller is a high-performance automated pneumatic pressure controller designed for professional calibration laboratories, production testing and automated pressure measurement applications. Combining fast pressure control, interchangeable reference pressure modules and advanced automation capabilities, the ADT 783-1K provides an efficient solution for pressure calibration from vacuum up to 1,000 psi / 70 bar.

The maximum pressure control range extends from:

-14.35 to 1,000 psi

or approximately:

-0.99 to 70 bar

The ADT 783-1K is primarily designed for:

Gauge Pressure

and can optionally support switching between:

Gauge and Absolute Pressure

when the appropriate barometric reference pressure module is installed.

Depending on the selected ADT151 pressure module, several measurement performance classes are available. Current configurations provide pressure accuracy specifications of:

0.02 % FS

0.01 % FS

or:

0.01 % RD

depending on the selected pressure module.

Pressure control stability is specified as:

<0.003 % FS

with typical performance of:

0.001 % FS

The ADT 783-1K also provides exceptionally fast pressure control, with a specified response time of:

<10 seconds

under the manufacturer’s defined test conditions.

Two interchangeable internal ADT151 pressure module bays allow the controller to cover multiple pressure ranges with one instrument. Automatic or fixed range switching can be selected according to the calibration task.

For professional calibration and metrological traceability of pressure controllers, reference modules and pressure measuring instruments, visit the europascal calibration service.

Your Benefits at a Glance

The Additel ADT 783-1K Pressure Controller provides:

  • Automated pneumatic pressure control
  • Pressure range from vacuum to 1,000 psi
  • Pressure range from approximately -0.99 to 70 bar
  • Gauge pressure control
  • Optional gauge/absolute pressure switching
  • Accuracy options down to 0.01 % FS
  • High-precision configurations down to 0.01 % of reading
  • Control stability <0.003 % FS
  • Typical control stability of 0.001 % FS
  • Pressure response time <10 seconds
  • Two interchangeable internal ADT151 pressure module bays
  • Automatic or fixed pressure range switching
  • Maximum high-to-low range ratio of 20:1
  • Fast, Standard and Custom control modes
  • Maximum overshoot <1 % FS
  • Maximum external load volume of 1,000 ml
  • External clean, dry air or nitrogen supply
  • Optional Contamination Prevention System
  • 7-inch color touchscreen
  • Automated calibration functions
  • Data logging
  • Remote communication interfaces
  • Suitable for laboratory, production and automated calibration systems

Pneumatic Pressure Control up to 70 bar

The Additel ADT 783-1K provides automated pneumatic pressure control over a range extending from approximately:

-0.99 to 70 bar

or:

-14.35 to 1,000 psi

This makes the controller suitable for a broad range of industrial and laboratory pressure calibration applications.

Unlike a portable pressure calibrator with an integrated pressure-generation pump, the ADT 783-1K is designed as a laboratory pressure controller and uses an:

external pneumatic pressure supply

The pressure medium should be clean, dry:

air or nitrogen

For optimum pressure control, the supply pressure should be appropriately matched to the maximum range of the installed internal control module.

External Pressure Supply

The ADT 783-1K requires an external gas supply for positive pressure generation.

The manufacturer specifies a maximum supply pressure of:

80 bar

for the ADT 783-1K configuration.

For optimum control performance, Additel recommends adjusting the source pressure to approximately 110 % of the maximum range of the internal pressure control module or 1 bar, whichever is greater.

This architecture is particularly suitable for stationary calibration laboratories because it combines a controlled external pressure supply with fast and highly stable electronic pressure regulation.

Gauge Pressure Control

The standard ADT 783-1K configuration is designed for:

Gauge Pressure

Gauge pressure is measured relative to atmospheric pressure:

p_gauge = p_absolute – p_atmosphere

The operating range includes negative gauge pressure as well as positive pressure up to 70 bar.

This allows the controller to cover pressure calibration tasks extending from vacuum relative to atmosphere through common industrial positive pressure ranges.

Optional Absolute Pressure Capability

The ADT 783-1K can optionally be configured for switching between:

Gauge Pressure

and:

Absolute Pressure

This requires the appropriate reference atmospheric pressure module.

When the barometric reference module is installed, the controller can use atmospheric pressure information to support absolute pressure measurements.

This provides additional flexibility for laboratories that need to calibrate both gauge and absolute pressure instruments.

Interchangeable ADT151 Pressure Modules

A major advantage of the Additel ADT 783-1K is its modular pressure measurement architecture.

The controller provides:

two interchangeable internal pressure module bays

for Additel ADT151 pressure modules.

The high-pressure module bay supports suitable ADT151 modules extending up to the 1,000 psi / 70 bar range.

The low-pressure bay can be equipped with a lower-range module to improve measurement performance at smaller pressures.

This modular architecture allows the controller to be adapted to changing calibration requirements without replacing the complete pressure controller.

Two Pressure Ranges in One Controller

A pressure calibration laboratory rarely works with only one pressure range.

For example, the same laboratory may need to calibrate:

  • 2 bar pressure transmitters
  • 10 bar digital pressure gauges
  • 25 bar pressure sensors
  • 60 bar pressure gauges

Using only a 70 bar reference sensor for every calibration may not provide the optimum measurement uncertainty at lower pressures.

The two-module architecture of the ADT 783-1K allows a lower-range reference module to be combined with the high-pressure module.

The controller can then select the most appropriate measurement range for the calibration task.

Maximum Range Ratio of 20:1

For the ADT 783-1K, Additel specifies a maximum high-to-low pressure range ratio of:

20:1

This allows the two installed pressure modules to cover substantially different ranges while maintaining suitable control performance.

The pressure range can be selected using:

Fixed

or:

Auto

range switching.

Automatic range selection can simplify calibration workflows involving devices with different pressure ranges.

Accuracy Options

The measurement performance of the Additel ADT 783-1K depends on the selected ADT151 pressure module.

Current configurations provide accuracy specifications of:

0.02 % FS

or:

0.01 % FS

with high-precision configurations available down to:

0.01 % RD

This allows the controller to be configured according to the uncertainty requirements of the calibration laboratory.

Precision and Accuracy

Additel distinguishes between:

Precision

and:

Accuracy

in the current ADT783 specifications.

Depending on the selected module, precision options for the ADT 783-1K include:

0.01 % FS

0.007 % FS

or:

0.008 % RD

The corresponding accuracy specifications are:

0.02 % FS

0.01 % FS

or:

0.01 % RD

Additel defines precision as including contributions such as linearity, hysteresis, repeatability, resolution and temperature compensation.

The accuracy specification additionally includes reference standard measurement uncertainty and annual drift.

This distinction is important when evaluating the ADT 783-1K for professional calibration applications.

Control Stability <0.003 % FS

The pressure control stability of the Additel ADT 783-1K is specified as:

<0.003 % FS

Under suitable conditions, Additel specifies typical performance of:

0.001 % FS

This high level of pressure stability is particularly valuable when calibrating high-resolution digital pressure gauges, pressure transmitters and reference pressure instruments.

Additional stabilization time may be required when the best 0.001 % FS performance is required.

Pressure Control in Less Than 10 Seconds

Speed is one of the key strengths of the ADT783 platform.

For the ADT 783-1K, Additel specifies a pressure control response time of:

<10 seconds

under the defined test conditions.

The manufacturer’s test uses an external load volume of 50 ml and a 20 % pressure step to reach 0.005 % FS stability.

This rapid control capability can significantly reduce the total time required for multipoint calibration procedures.

Three Pressure Control Modes

The ADT 783-1K provides three control modes:

Fast

Standard

and:

Custom

These modes allow the pressure control behavior to be adapted to the calibration application.

Fast control is useful when short calibration cycle times are important, while other settings can be selected when the connected device or pressure volume requires different control characteristics.

Maximum Overshoot <1 % FS

Pressure overshoot can be important when calibrating sensitive instruments.

The ADT 783-1K specifies a maximum pressure overshoot of:

<1 % FS

This controlled approach to the target pressure helps protect the device under test and supports efficient automated calibration.

Load Volumes up to 1,000 ml

The ADT 783-1K supports an external load volume of up to:

1,000 ml

This provides flexibility when connecting different devices under test, manifolds, tubing systems and calibration setups.

The actual pressure response time depends on factors including:

  • Connected volume
  • Pressure step
  • Supply pressure
  • Selected control mode
  • Target stability
  • Pressure range

Pressure Port

The ADT 783-1K uses a:

1/4 BSP male

pressure connection.

Pressure ports incorporate filtration to help protect the controller from contamination.

Additel specifies filters in the approximately:

40 to 100 µm

range for the pressure ports.

Optional Contamination Prevention System

An optional:

Contamination Prevention System (CPS)

is available for the ADT 783-1K.

Contamination protection is particularly useful when devices under test may contain particles, moisture, oil or other contamination that could enter the pressure controller.

Protecting the reference modules and internal pressure-control system helps maintain reliable operation and metrological performance.

Fast Module Changes

The ADT783 is designed around Additel’s interchangeable ADT151 pressure modules.

Modules can be exchanged when calibration workloads change.

This modular approach allows one pressure controller to support several pressure ranges without permanently integrating a dedicated reference sensor for every application.

For calibration laboratories with changing workloads, this can improve equipment utilization and reduce the need for multiple complete controllers.

Automatic Calibration Applications

The Additel ADT 783-1K is particularly well suited to automated calibration.

The pressure controller can generate and regulate defined pressure points while measurement results are recorded through an appropriate calibration workflow.

A typical calibration sequence may include:

  1. Zero pressure
  2. Increasing pressure points
  3. Full-scale pressure
  4. Decreasing pressure points
  5. Return to zero

This allows characteristics such as indication error, repeatability and hysteresis of the device under test to be evaluated efficiently.

Digital Pressure Gauge Calibration

The ADT 783-1K is ideal for calibrating digital pressure gauges.

The controller automatically generates a stable reference pressure while the indication of the gauge is compared with the reference value.

Its rapid pressure control can significantly shorten multipoint calibrations compared with manual pressure generation.

Pressure Transmitter Calibration

Pressure transmitters can also be calibrated using the ADT 783-1K as the pneumatic reference pressure source.

Typical devices include:

  • Gauge pressure transmitters
  • Absolute pressure transmitters with the appropriate controller configuration
  • Smart pressure transmitters
  • Industrial pressure transducers

The controller’s automation and remote communication capabilities make it particularly useful in laboratory and production test systems.

Pressure Sensor Production Testing

The fast response and modular pressure ranges also make the ADT 783-1K suitable for production environments.

Pressure sensors can be tested at predefined pressure points as part of an automated production sequence.

The combination of:

  • Fast control
  • Stable pressure
  • Interchangeable modules
  • Remote communication
  • Automated operation

provides a strong platform for repetitive pressure testing.

Communication and Remote Operation

The ADT783 platform provides several interfaces for integration into automated calibration systems.

Available communication options include interfaces such as:

  • LAN / Ethernet
  • USB
  • RS232

These interfaces allow the pressure controller to communicate with computers, calibration software and automated test systems.

The controller also supports programming commands for integration into custom calibration applications.

Modbus Support

Additel provides:

Modbus commands

for the ADT773, ADT783 and ADT793 pressure controller platform.

This allows the ADT 783-1K to be integrated into compatible industrial automation and laboratory control systems.

For automated test benches, Modbus communication can provide a convenient method of remotely setting pressures and reading instrument information.

7-Inch Color Touchscreen

The ADT783 uses a large:

7-inch color touchscreen

with an icon-based user interface.

Pressure setpoints, control modes, measurement values and configuration parameters can be accessed directly from the display.

The touchscreen architecture makes the controller easy to operate while still providing the advanced functionality required in professional calibration laboratories.

Technical Highlights

Model: Additel ADT 783-1K

Device type: Automated Pneumatic Pressure Controller

Maximum pressure: 1,000 psi

Maximum pressure: 70 bar

Minimum pressure: -14.35 psi

Minimum pressure: -0.99 bar

Standard pressure type: Gauge

Gauge/absolute switching: Optional with barometric reference module

Accuracy: 0.02 % FS, 0.01 % FS or 0.01 % RD depending on module

Precision: 0.01 % FS, 0.007 % FS or 0.008 % RD depending on module

Control stability: <0.003 % FS

Typical control stability: 0.001 % FS

Control response time: <10 seconds under specified conditions

Internal pressure module bays: 2

Pressure modules: Additel ADT151

Maximum high/low range ratio: 20:1

Range switching: Fixed or Auto

Control modes: Fast, Standard, Custom

Maximum overshoot: <1 % FS

Maximum load volume: 1,000 ml

Pressure medium: Clean, dry air or nitrogen

Pressure supply: External

Maximum supply pressure: 80 bar

Pressure connection: 1/4 BSP male

Contamination Prevention System: Optional

Display: 7-inch color touchscreen

Remote communication: Yes

Modbus: Supported

Technical Data

Feature Additel ADT 783-1K
Device type Automated Pneumatic Pressure Controller
Control range -14.35 to 1,000 psi
Control range -0.99 to 70 bar
Standard pressure type Gauge
Absolute pressure Optional with barometric reference
Accuracy 0.02 % FS / 0.01 % FS / 0.01 % RD
Precision 0.01 % FS / 0.007 % FS / 0.008 % RD
Control stability <0.003 % FS
Typical control stability 0.001 % FS
Response time <10 s
Internal module bays 2
Pressure modules ADT151
Maximum range ratio 20:1
Range selection Fixed / Auto
Control modes Fast / Standard / Custom
Maximum overshoot <1 % FS
Maximum load volume 1,000 ml
Pressure medium Clean, dry air or nitrogen
Pressure supply External
Maximum supply pressure 80 bar
Pressure port 1/4 BSP male
CPS Optional
Remote control Yes
Modbus Supported

ADT 783-1K vs. ADT 761A-1K

Although the ADT 783-1K and ADT 761A-1K both cover pressures up to approximately 70 bar, they are designed for different applications.

The ADT 783-1K is primarily a:

Benchtop Pressure Controller

while the ADT 761A-1K is a:

Portable Automatic Pressure Calibrator

The ADT 761A-1K incorporates its own electric pressure-generation system and is designed for mobile field calibration.

The ADT 783-1K instead uses an external supply of clean, dry air or nitrogen and is optimized for rapid, stable pressure control in calibration laboratories, production testing and automated calibration systems.

The ADT 783-1K provides control stability of less than 0.003 % FS, with typical performance of 0.001 % FS, and can reach the specified pressure control condition in less than 10 seconds under Additel’s defined test conditions.

For stationary laboratory applications requiring rapid automated pressure control, the ADT 783-1K is therefore the more specialized solution.

ADT 783 Series Overview

The Additel ADT783 series covers several pneumatic pressure ranges.

Model Maximum Range Pressure Type
ADT 783-D 35 psi / 2.5 bar Differential
ADT 783-1K 1,000 psi / 70 bar Gauge
ADT 783-3.6K 3,600 psi / 250 bar Gauge
ADT 783-6K 6,000 psi / 420 bar Gauge

The ADT 783-1K is particularly suitable for laboratories requiring automatic pneumatic pressure control from vacuum through common industrial pressure ranges up to 70 bar.

Higher-pressure applications can be covered by the ADT 783-3.6K and ADT 783-6K configurations.

Typical Applications

The Additel ADT 783-1K Pressure Controller is suitable for:

  • Digital pressure gauge calibration
  • Analog pressure gauge calibration
  • Pressure transmitter calibration
  • Pressure sensor calibration
  • Pressure transducer calibration
  • Reference pressure instrument calibration
  • Automated calibration systems
  • Production pressure testing
  • Research and development
  • Quality assurance
  • Pressure laboratory automation
  • Multipoint pressure calibration
  • Gauge pressure calibration
  • Optional absolute pressure calibration

Typical users include:

  • Professional calibration laboratories
  • ISO/IEC 17025 calibration laboratories
  • Industrial calibration departments
  • Instrument manufacturers
  • Pressure sensor manufacturers
  • Quality assurance departments
  • Research laboratories
  • Production test facilities
  • Automated test system operators

Calibration and Service by europascal

A high-performance pressure controller depends on reliable reference pressure measurement.

The interchangeable ADT151 pressure modules should therefore be included in the calibration strategy according to the ranges actually used.

For systems equipped with an optional barometric reference module, this reference should also be considered when absolute pressure functionality is required.

The europascal calibration service provides professional calibration solutions for pressure controllers, reference pressure modules, digital pressure gauges, pressure transmitters and pressure sensors.

europascal can also support customers in selecting suitable ADT151 modules, pressure ranges, accessories and system configurations for the ADT 783-1K.

Why Choose the Additel ADT 783-1K?

The Additel ADT 783-1K Precise Pressure Controller combines automated pneumatic pressure control from approximately -0.99 to 70 bar, interchangeable reference pressure modules and exceptional control performance in a modern laboratory platform.

Depending on the selected ADT151 pressure module, accuracy options extend from 0.02 % FS through 0.01 % FS to high-precision configurations reaching 0.01 % of reading.

Control stability is specified at <0.003 % FS, with typical performance of 0.001 % FS. Pressure control can be achieved in less than 10 seconds under the manufacturer’s defined test conditions.

Two interchangeable pressure module bays, automatic range switching, optional gauge-to-absolute pressure conversion, remote communication and Modbus support make the ADT 783-1K particularly suitable for professional calibration laboratories and automated pressure test systems.

Looking for a fast and precise automated pressure controller up to 70 bar? Contact europascal for expert advice on the Additel ADT 783-1K, suitable ADT151 pressure modules, accessories and professional calibration services.

Further Additel pressure controllers and calibrators from europascal can be found here.

Professional pressure calibration services from europascal can be found here.

Manufacturer information about the Additel ADT 783 can be found on the official Additel website.

Download Datasheet

Medium

Druckbereich

Genauigkeit

Datenlogger

Schnittstelle

Strom-versorgung

Manometer Typen

HART-Kommunikation

Bauart

Kanäle

Other Products

Scroll to Top

Request a Quote Now!

Download Product Datasheet

Enter your email address below to access the official product datasheet, including technical specifications, measurement ranges, product features, and performance details.