The europascal BH2-4000B High Pressure Deadweight Tester is a high-performance hydraulic pressure standard designed for demanding calibration applications from 40 to 4,000 bar. With a standard accuracy of 0.02 % of reading and enhanced performance up to 0.008 % on request, the BH2-4000B provides precise physical pressure generation for professional high-pressure calibration laboratories.
The instrument covers a measuring range from 40 to 4,000 bar, corresponding to 4 to 400 MPa, 4,000 to 400,000 kPa and 600 to 60,000 psi according to the manufacturer.
The BH2-4000B uses a single precision piston-cylinder system with a typical piston area of 1.96315 mm². Both the piston and cylinder are manufactured from tungsten carbide, providing the mechanical properties required for reliable pressure generation at pressures up to 400 MPa.
The hydraulic pressure medium is sebacate oil with a specified density of approximately 915 kg/m³. Pressure is generated using the CH2-4000B generator base.
The reference pressure is physically generated by calibrated masses acting on the precision piston-cylinder assembly. The robust BH2 platform can additionally be equipped with optional motorization, piston position visualization and automatic pressure calculation for more efficient calibration workflows.
The europascal calibration service provides professional calibration solutions for deadweight testers, pressure balances, high-pressure gauges, pressure sensors, transmitters and other pressure measuring equipment.
Your Benefits at a Glance
The europascal BH2-4000B High Pressure Deadweight Tester provides:
- Hydraulic high-pressure deadweight tester
- Measuring range from 40 to 4,000 bar
- Measuring range from 4 to 400 MPa
- Measuring range from 4,000 to 400,000 kPa
- Manufacturer range from 600 to 60,000 psi
- Standard accuracy of 0.02 % of reading
- Enhanced accuracy up to 0.008 % on request
- Single-piston configuration
- Typical piston area of 1.96315 mm²
- Tungsten carbide piston
- Tungsten carbide cylinder
- CH2-4000B generator base
- Sebacate oil as pressure medium
- Fluid density approximately 915 kg/m³
- Instrument weight approximately 22 kg without masses
- Stainless steel calibration masses
- Standard or customer-specific gravity
- Optional motorized mass rotation
- Optional piston position visualization
- Optional automatic pressure calculation
- Optional computer data transfer
Measuring Range from 40 to 4,000 bar
The europascal BH2-4000B covers a hydraulic pressure range from:
40 to 4,000 bar
This corresponds to:
4 to 400 MPa
and:
4,000 to 400,000 kPa
The manufacturer specifies:
600 to 60,000 psi
The maximum pressure is therefore:
4,000 bar / 400 MPa
This wide range makes the BH2-4000B particularly suitable for laboratories requiring a physical reference standard for calibration of high-pressure measuring equipment.
Accuracy of 0.02 % of Reading
The standard accuracy of the europascal BH2-4000B is:
0.02 % of reading
For applications with more demanding metrological requirements, AREMECA specifies an enhanced accuracy:
up to 0.008 % on request
This allows the instrument configuration to be selected according to the required calibration capability.
The actual measurement uncertainty of a calibration additionally depends on the individual calibration data, operating conditions and complete measurement uncertainty budget.
Physical Piston Gauge Measurement Principle
The europascal BH2-4000B High Pressure Deadweight Tester operates according to the fundamental physical principle of a piston gauge.
The basic relationship is:
p = F / A
where:
p = generated pressure
F = force acting on the piston
A = effective piston area
The gravitational force generated by the calibration masses is primarily determined by:
F = m × g
where:
m = applied mass
g = gravitational acceleration
The simplified pressure relationship is therefore:
p = (m × g) / A
For precision pressure calibration, additional correction quantities are considered according to the calibration data and operating conditions.
Typical Piston Area of 1.96315 mm²
AREMECA specifies a typical piston area for the BH2-4000B of:
1.96315 mm²
This relatively small piston area makes it possible to generate very high pressures with manageable calibration masses.
The relationship can be seen directly from:
p = F / A
For a given force, reducing the effective piston area increases the generated pressure.
For precise calibration, the individually calibrated effective area of the piston-cylinder system is used rather than simply relying on the nominal or typical area.
Tungsten Carbide Piston and Cylinder
The BH2-4000B uses:
Piston: tungsten carbide
Cylinder: tungsten carbide
Tungsten carbide provides excellent hardness, wear resistance and dimensional stability.
These characteristics are particularly valuable for piston-cylinder systems operating at very high hydraulic pressures.
At a maximum pressure of:
4,000 bar / 400 MPa
the piston-cylinder assembly is exposed to substantial mechanical loading.
The material selection therefore contributes to stable and reproducible operation of the high-pressure reference.
Single-Piston Configuration
The europascal BH2-4000B incorporates:
one piston
and:
one precision piston-cylinder system
The complete measuring range from 40 to 4,000 bar is generated with this single piston system.
This provides straightforward operation because no change between separate low- and high-pressure piston assemblies is required within the specified range.
Sebacate Oil as Pressure Medium
The BH2-4000B operates with:
sebacate oil
AREMECA specifies a fluid density of approximately:
915 kg/m³
The hydraulic pressure medium is an important part of the complete pressure calibration system.
Correct filling, cleanliness and venting are essential for stable operation.
Trapped gas can influence settling behavior and pressure stability, while contamination can adversely affect precision hydraulic components.
Compatibility between the pressure medium and the connected device under test should always be verified.
CH2-4000B Generator Base
The europascal BH2-4000B uses the:
CH2-4000B generator base
The pressure generator allows the operator to generate the hydraulic pressure required to raise the piston and balance the applied calibration masses.
The mechanical pressure generator provides a robust solution for laboratory use and forms an integral part of the complete high-pressure deadweight tester.
High-Pressure Calibration up to 400 MPa
A maximum pressure of 4,000 bar corresponds to:
400 MPa
At these pressure levels, every component in the hydraulic calibration system must be suitable for the maximum applied pressure.
This includes:
- High-pressure tubing
- Fittings
- Adapters
- Valves
- Seals
- Pressure generator
- Device under test
- Pressure relief equipment
Correct high-pressure operating procedures are essential.
The system should be depressurized completely before hydraulic connections are changed.
Measurement Uncertainty Contributions
The complete measurement uncertainty of a high-pressure calibration depends on several quantities.
Relevant contributions can include:
- Calibration uncertainty of the masses
- Effective piston area
- Pressure coefficient
- Elastic deformation of the piston-cylinder system
- Local gravitational acceleration
- Air buoyancy
- Piston-cylinder temperature
- Thermal expansion
- Hydraulic fluid properties
- Hydrostatic head difference
- Repeatability
- Sensitivity
- Pressure stability
- Resolution of the device under test
- Environmental conditions
At several thousand bar, pressure-related deformation of the piston-cylinder system is particularly relevant.
The effective area under operating pressure should therefore be calculated using the individual calibration parameters of the piston-cylinder assembly.
Stainless Steel Calibration Masses
The BH2 series uses stainless steel calibration masses.
These masses generate the defined gravitational force acting on the precision piston.
Together with the effective piston area and gravitational acceleration, the calibrated mass values determine the generated reference pressure.
The mass set can be configured in different pressure units according to the application.
Standard or Local Gravity
The force generated by a calibration mass depends directly on gravitational acceleration.
The conventional standard value is:
g = 9.80665 m/s²
Actual gravitational acceleration varies according to geographical location.
The BH2-4000B can therefore be configured for:
standard gravity
or:
customer-specific local gravity
For high-accuracy calibration, the correct gravity value is important because a relative difference in gravity produces a corresponding relative difference in the force generated by the calibration masses.
Operating Temperature
AREMECA specifies an operating temperature range of:
18 to 28 °C
Stable environmental conditions are recommended for high-accuracy pressure calibration.
Temperature is particularly relevant because it influences the dimensions and therefore the effective area of the piston-cylinder assembly.
The actual piston temperature and relevant correction data should be considered according to the calibration procedure.
Instrument Weight
The standard BH2-4000B weighs approximately:
22 kg without calibration masses
This provides a relatively compact reference instrument considering its maximum pressure capability of 400 MPa.
The calibration masses add additional weight depending on the selected pressure unit and mass-set configuration.
Optional OP0004 Motorization
The europascal BH2-4000B can be equipped with:
Option OP0004
This provides:
- Motorized rotation of the calibration masses
- Piston position visualization using indicator lights
Rotation of the piston helps minimize friction effects between the piston and cylinder.
Motorized operation reduces manual intervention and provides additional convenience during repetitive calibration procedures.
Optional OP0301 Calculator Module
For more advanced calibration workflows, the BH2-4000B can be equipped with:
Option OP0301
This adds functions including:
- Motorized operation
- Dynamic piston position visualization
- Automatic pressure calculation
- Automatic pressure display
- Consideration of temperature
- Consideration of atmospheric pressure
- Consideration of relative humidity
- Computer data transfer
- Calibration data traceability
This allows the standard BH2 platform to be expanded toward a more automated calibration system.
Scalable BH2 Platform
A major advantage of the BH2 series is its scalable design.
The standard BH2-4000B provides a robust manually operated high-performance pressure balance.
Depending on laboratory requirements, additional functionality can be added.
The system can therefore be configured for:
- Conventional manual operation
- Motorized mass rotation
- Piston position visualization
- Automatic pressure calculation
- Digital data transfer
This makes the BH2-4000B suitable for both traditional pressure metrology and more automated calibration workflows.
Technical Highlights
Model: europascal BH2-4000B
Device type: High Pressure Hydraulic Deadweight Tester
Configuration: Single piston
Number of pistons: 1
Measuring range: 40 to 4,000 bar
Measuring range: 4 to 400 MPa
Measuring range: 4,000 to 400,000 kPa
Manufacturer imperial range: 600 to 60,000 psi
Maximum pressure: 4,000 bar / 400 MPa
Standard accuracy: 0.02 % of reading
Enhanced accuracy: Up to 0.008 % on request
Typical piston area: 1.96315 mm²
Piston material: Tungsten carbide
Cylinder material: Tungsten carbide
Generator base: CH2-4000B
Pressure medium: Sebacate oil
Fluid density: Approx. 915 kg/m³
Instrument weight without masses: Approx. 22 kg
Measurement principle: Piston gauge / deadweight tester
Operating temperature: 18 to 28 °C
Standard gravity: 9.80665 m/s²
Local gravity: Available
Motorization: Optional OP0004 / OP0301
Automatic pressure calculation: Optional OP0301
Computer data transfer: Optional OP0301
Technical Data
| Feature | europascal BH2-4000B |
|---|---|
| Manufacturer | europascal Aremeca |
| Model | BH2-4000B |
| Device type | High Pressure Hydraulic Deadweight Tester |
| Configuration | Single piston |
| Number of pistons | 1 |
| Measuring range | 40–4,000 bar |
| Measuring range | 4–400 MPa |
| Measuring range | 4,000–400,000 kPa |
| Manufacturer imperial range | 600–60,000 psi |
| Maximum pressure | 4,000 bar / 400 MPa |
| Standard accuracy | 0.02 % of reading |
| Enhanced accuracy | Up to 0.008 % on request |
| Typical piston area | 1.96315 mm² |
| Piston material | Tungsten carbide |
| Cylinder material | Tungsten carbide |
| Generator base | CH2-4000B |
| Pressure medium | Sebacate oil |
| Fluid density | Approx. 915 kg/m³ |
| Weight without masses | Approx. 22 kg |
| Operating temperature | 18–28 °C |
| Standard gravity | 9.80665 m/s² |
| Motorization | Optional |
| Piston position visualization | Optional |
| Automatic pressure calculation | Optional OP0301 |
| Computer data transfer | Optional OP0301 |
Position within the BH2 High-Pressure Series
The BH2 series provides several high-pressure configurations for different maximum pressures.
| Model | Pressure Range | Accuracy |
|---|---|---|
| BH2-2000B | 40–2,000 bar | 0.02 % |
| BH2-2500B | 40–2,500 bar | 0.02 % |
| BH2-3000B | 40–3,000 bar | 0.02 % |
| BH2-4000B | 40–4,000 bar | 0.02 % |
| BH2-5000B | 40–5,000 bar | 0.02 % |
| BH2-6000B | 200–6,000 bar | 0.05 % |
| BH2-7000B | 200–7,000 bar | 0.05 % |
The BH2-4000B therefore provides a particularly wide high-pressure range while retaining a lower limit of only 40 bar.
BH2-4000B vs. BH2-5000B
The BH2-4000B and BH2-5000B are closely related high-pressure pressure balances.
| Feature | BH2-4000B | BH2-5000B |
|---|---|---|
| Minimum pressure | 40 bar | 40 bar |
| Maximum pressure | 4,000 bar | 5,000 bar |
| Maximum pressure | 400 MPa | 500 MPa |
| Number of pistons | 1 | 1 |
| Typical piston area | 1.96315 mm² | 1.96315 mm² |
| Piston material | Tungsten carbide | Tungsten carbide |
| Cylinder material | Tungsten carbide | Tungsten carbide |
| Pressure medium | Sebacate oil | Sebacate oil |
| Standard operation | Manual | Manual |
| Motorization | Optional | Optional |
| Automatic calculation | Optional | Optional |
The BH2-4000B is therefore an excellent choice when calibration requirements do not exceed 400 MPa, while the BH2-5000B extends the maximum pressure to 500 MPa.
Typical Calibration Applications
The europascal BH2-4000B High Pressure Deadweight Tester is suitable for testing, adjustment and calibration of:
- High-pressure gauges
- Digital pressure gauges
- High-pressure sensors
- Pressure transducers
- Pressure transmitters
- Pressure switches
- Reference pressure instruments
- Hydraulic high-pressure systems
- Industrial pressure measurement equipment
Typical users include:
- Professional calibration laboratories
- Pressure instrument manufacturers
- Industrial quality assurance departments
- Research laboratories
- High-pressure technology facilities
- Calibration service organizations
Calibration and Traceability
Regular calibration of a high-pressure deadweight tester is important for maintaining reliable metrological performance.
Important parameters include:
- Effective piston area
- Pressure coefficient
- Piston-cylinder characteristics
- Calibration mass values
- Local gravitational acceleration
- Temperature
- Repeatability
- Relevant correction quantities
AREMECA supplies the instrument with a factory calibration certificate traceable to national standards. An accredited COFRAC or DKD/DAkkS calibration certificate is available as an option.
The europascal calibration service provides professional calibration solutions for pressure balances, deadweight testers, high-pressure gauges, pressure sensors and pressure transmitters.
europascal can also provide support with maintenance, configuration and selection of suitable high-pressure calibration equipment.
Why Choose the europascal BH2-4000B?
The europascal BH2-4000B Precise High Pressure Deadweight Tester combines a wide hydraulic pressure range from 40 to 4,000 bar with a standard accuracy of 0.02 % of reading and enhanced accuracy up to 0.008 % on request.
Its single-piston design, typical piston area of 1.96315 mm², tungsten carbide piston and cylinder, sebacate oil pressure medium and CH2-4000B pressure generator provide a robust physical reference for demanding calibration applications up to 400 MPa.
The standard BH2 platform provides reliable manual operation, while optional motorization, piston position visualization, automatic pressure calculation and computer data transfer allow the system to be adapted to more automated calibration workflows.
For professional laboratories requiring a high-pressure reference with a particularly broad measuring range, the BH2-4000B provides an excellent combination of pressure capability, metrological performance and configurable operation.
Looking for the europascal BH2-4000B or a hydraulic high-pressure calibration solution up to 4,000 bar? Contact europascal for expert advice, system configuration and professional calibration services tailored to your requirements.