The Additel AccuMac AM 1760 is a high-precision secondary standard platinum resistance thermometer for professional temperature measurement and calibration tasks. This reference probe was developed specifically for daily use in demanding calibration laboratories and combines high accuracy with excellent stability, mechanical robustness and a wide usable temperature range.
With an operating range of −200 °C to 670 °C, the Additel AccuMac AM 1760 covers a very large part of contact thermometry. The probe is therefore suitable both for low-temperature measurements and for comparison calibrations at high temperatures. Additel explicitly describes the AM1760 series as a reliable secondary standard thermometer for everyday laboratory use.
The sensing element has a nominal resistance of 100 Ω at 0.01 °C and a temperature coefficient of 0.003925 Ω/Ω/°C. This distinguishes the AM 1760 in particular from the related AM 1762 with 25 Ω. The 100 Ω version is especially interesting for measurement systems and temperature readouts optimized for high-quality Pt100-type reference probes.
The sensing element is designed to protect the platinum wire from contamination as well as possible, even at high temperatures. Additel uses a specially designed support structure and filling materials to achieve a balance between low hysteresis, mechanical robustness and thermal shock resistance.
At europascal you receive professional solutions for temperature measurement and temperature calibration. You can find further suitable products in the temperature sensors section at europascal. For the regular traceable calibration of your temperature probes and reference thermometers, the europascal Calibration Service is also at your disposal.
Your Benefits at a Glance
- Secondary standard platinum resistance thermometer
- Temperature range from −200 °C to 670 °C
- Nominal resistance of 100 Ω at 0.01 °C
- Temperature coefficient 0.003925 Ω/Ω/°C
- Accuracy ±0.010 °C at −196 °C
- Accuracy ±0.006 °C at 0.01 °C
- Accuracy ±0.015 °C at 420 °C
- Accuracy ±0.025 °C at 661 °C
- Short-term stability less than 0.002 °C
- Drift less than 0.004 °C at the triple point of water after 100 hours at 661 °C
- Thermal shock change less than 0.002 °C after ten temperature cycles
- Self-heating approx. 0.0015 °C at 0.5 mA
- Four-wire connection
- Measuring current 0.5 mA
- Inconel protective sheath
- Sensing element approx. 42 mm long
- Sensor position approx. 5 mm from the tip
- Insulation resistance greater than 1,000 MΩ at room temperature
- Two probe lengths available
- 305 mm version
- 508 mm version
- Suitable for fixed-point and comparison calibrations
- Meets ITS-90 criteria for reference thermometers
- ISO 17025 accredited calibration according to the manufacturer
The combination of high accuracy, very low drift and robust mechanical design makes the Additel AccuMac AM 1760 a powerful reference probe for professional temperature calibration.
Additel AccuMac AM 1760 for Demanding Calibration Applications
The Additel AccuMac AM 1760 was developed for applications that require a very stable and reproducible reference temperature.
Typical fields of application include:
- Calibration laboratories
- Accredited calibration laboratories
- In-house calibration laboratories
- Research and development
- Temperature metrology
- Comparison calibrations
- Fixed-point calibrations
- Calibration baths
- Liquid baths
- Temperature furnaces
- Dry-block calibrators
- Test equipment management
- Characterization of temperature sources
In a classic comparison calibration, the AM 1760 is placed together with the device under test in a temperature source that is as stable and homogeneous as possible.
The resistance of the reference probe is measured with a suitable precision temperature readout or a resistance bridge. The individual calibration coefficients then allow the reference temperature to be calculated precisely.
Temperature Range from −200 °C to 670 °C
The wide temperature range is one of the most important features of the Additel AccuMac AM 1760.
The probe can be used over a range of:
−200 °C to 670 °C
A single reference probe therefore covers numerous typical temperature calibration applications.
In the low-temperature range, the sensor is suitable, for example, for measurements close to liquid nitrogen. In the medium temperature range, it can be used in high-quality liquid baths. For higher temperatures, suitable calibration furnaces or fixed-point cells are typical applications.
High Accuracy of the Additel AccuMac AM 1760
The Additel AccuMac AM 1760 offers a very high specified accuracy.
Additel specifies the following values:
| Temperature | Calibrated accuracy k = 2 |
|---|---|
| −196 °C | ±0.010 °C |
| 0.01 °C | ±0.006 °C |
| 420 °C | ±0.015 °C |
| 661 °C | ±0.025 °C |
These values include both the calibration and the contribution of the 100-hour drift.
The accuracy of ±0.006 °C at the triple point of water in particular makes the AM 1760 attractive for demanding reference measurements.
The actual measurement uncertainty of a complete system also depends on the temperature readout used, the resistance measurement, the calibration source, the immersion depth, self-heating and other influencing factors.
Excellent Short-Term Stability
For a secondary reference thermometer, absolute accuracy is not the only decisive factor.
The short-term stability of the resistance value is just as important.
Additel specifies a short-term stability for the AccuMac AM 1760 of:
less than 0.002 °C
This high stability supports reproducible measurements and is particularly advantageous for comparison calibrations.
Low Long-Term Drift
Long-term stability also plays a central role for a reference probe.
For the Additel AccuMac AM 1760, Additel specifies a drift at the triple point of water of:
less than 0.004 °C after 100 hours at 661 °C
This shows that the probe was specifically designed for high stability even after prolonged thermal stress.
Low drift reduces the influence of long-term sensor changes on the uncertainty budget and can be taken into account when defining suitable recalibration intervals.
Good Thermal Shock Resistance
Temperature probes can be subjected to high mechanical and thermal stress during rapid changes between very low and high temperatures.
The Additel AccuMac AM 1760 has therefore also been optimized for thermal shock resistance.
Additel specifies a change of:
less than 0.002 °C after ten complete temperature cycles between minimum and maximum temperature
This supports high reliability in applications in which the probe is regularly moved between different temperature ranges.
100 Ω Nominal Resistance
The Additel AccuMac AM 1760 has a nominal resistance of:
100 Ω at 0.01 °C
This version is particularly interesting for high-quality resistance measuring instruments and temperature readouts designed for 100 Ω reference probes.
The temperature coefficient is:
0.003925 Ω/Ω/°C
This distinguishes the probe from typical industrial Pt100 sensors according to IEC 60751 with an alpha value of approx. 0.00385 Ω/Ω/°C.
The AM 1760 should therefore not be regarded as an ordinary industrial Pt100, but as a high-quality reference PRT with a characteristic curve optimized for metrological applications.
Low Self-Heating
With resistance thermometers, the required measuring current causes a small electrical power dissipation.
This can slightly heat the sensing element.
For the Additel AccuMac AM 1760, Additel specifies a self-heating of approximately:
0.0015 °C
at a measuring current of:
0.5 mA
The influence of self-heating therefore remains small and can be taken into account in a controlled manner in high-precision measurements.
Four-Wire Technology
The Additel AccuMac AM 1760 is operated in four-wire configuration.
Two wires supply the measuring current and two further wires measure the voltage across the sensing element.
This largely eliminates the influence of the lead wires on the resistance measurement.
In demanding temperature calibrations in particular, four-wire technology is essential for achieving reproducible resistance values with low measurement uncertainty.
Robust Inconel Protective Sheath
The protective sheath of the Additel AccuMac AM 1760 is made of Inconel.
This material offers high temperature resistance and good mechanical properties.
The outer diameter of the probe is:
6.35 mm
The probe thus offers a good compromise between mechanical stability, sufficient rigidity and good thermal behavior.
Two Available Probe Lengths
The Additel AccuMac AM 1760 is available in two straight versions:
AM1760-12
Length approx. 305 mm
and
AM1760-20
Length approx. 508 mm
The longer version is particularly interesting for temperature sources that require a greater immersion depth.
These can include deep liquid baths, fixed-point cells or high-temperature furnaces.
Sensor Position and Sensor Length
The actual sensing element of the AM 1760 has a length of approximately:
42 mm
The active sensor zone begins approximately:
5 mm from the probe tip
This clearly defined position makes it easier to place the sensor correctly within a temperature source.
In high-precision comparison calibrations, care should nevertheless be taken to ensure sufficient immersion depth so that heat conduction along the probe stem remains as low as possible.
Response Behavior
For the AM1760 and AM1762 series, a typical 63 % response time of approximately:
9 seconds
is specified at a defined water flow of 1 m/s.
However, the actual settling time depends strongly on the temperature source used.
In a well-stirred liquid bath, for example, the thermal behavior is considerably more favorable than in still air.
For high-precision calibrations, always wait until both the reference and the device under test have reached a stable thermal equilibrium.
Technical Highlights
The Additel AccuMac AM 1760 combines high accuracy with very good stability:
- Temperature range −200 °C to 670 °C
- Nominal 100 Ω at 0.01 °C
- Temperature coefficient 0.003925 Ω/Ω/°C
- Accuracy up to ±0.006 °C
- Short-term stability less than 0.002 °C
- Drift less than 0.004 °C after 100 hours at 661 °C
- Thermal shock change less than 0.002 °C
- Self-heating approx. 0.0015 °C at 0.5 mA
- Four-wire technology
- Inconel sheath
- 6.35 mm outer diameter
- Sensor length approx. 42 mm
- Sensor position approx. 5 mm from the tip
- Insulation resistance greater than 1,000 MΩ
- 305 mm or 508 mm length
- ITS-90 criteria met
- ISO 17025 accredited calibration
Technical Data
| Feature | Additel AccuMac AM 1760 |
|---|---|
| Manufacturer | Additel AccuMac |
| Model | AM 1760 |
| Device type | Secondary standard platinum resistance thermometer |
| Sensor type | Secondary SPRT |
| Temperature range | −200 °C to 670 °C |
| Nominal resistance at 0.01 °C | 100 Ω |
| Temperature coefficient | 0.003925 Ω/Ω/°C |
| Accuracy at −196 °C | ±0.010 °C |
| Accuracy at 0.01 °C | ±0.006 °C |
| Accuracy at 420 °C | ±0.015 °C |
| Accuracy at 661 °C | ±0.025 °C |
| Short-term stability | < 0.002 °C |
| Long-term drift | < 0.004 °C at TPW after 100 h at 661 °C |
| Thermal shock | < 0.002 °C after 10 cycles |
| Self-heating | approx. 0.0015 °C at 0.5 mA |
| Measuring current | 0.5 mA |
| Sensor length | approx. 42 mm |
| Sensor position | approx. 5 mm from the tip |
| Insulation resistance | > 1,000 MΩ at room temperature |
| Sheath material | Inconel |
| Outer diameter | 6.35 mm |
| Probe lengths | 305 mm or 508 mm |
| Connection | Four-wire |
| Lead wire | Teflon-insulated copper wire |
| Handle temperature range | −50 °C to 180 °C |
| Calibration | ISO 17025 accredited according to the manufacturer |
| Temperature scale | Meets ITS-90 criteria |
Additel AccuMac AM 1760 Compared with the AM 1762
The Additel AccuMac AM 1760 and the Additel AccuMac AM 1762 belong to the same series of secondary standard platinum resistance thermometers.
Both models cover a temperature range of:
−200 °C to 670 °C
and offer comparably high stability and accuracy.
The decisive difference lies in the nominal resistance.
The Additel AccuMac AM 1760 has:
100 Ω at 0.01 °C
The Additel AccuMac AM 1762, on the other hand, has:
25 Ω at 0.01 °C
The appropriate version can thus be selected according to the reference readout, resistance bridge and reference resistor used.
The AM 1760 is particularly suitable for high-quality measurement systems designed for 100 Ω reference PRTs.
The AM 1762, by contrast, is particularly interesting for systems with 25 Ω reference PRTs.
Why Choose the Additel AccuMac AM 1760?
The Additel AccuMac AM 1760 offers a very well-balanced combination of high accuracy, stability and robustness.
The temperature range of −200 °C to 670 °C enables numerous calibration tasks with a single reference probe.
The accuracy of ±0.006 °C at the triple point of water, the short-term stability of less than 0.002 °C and the very low long-term drift support demanding uncertainty budgets.
The 100 Ω nominal resistance makes the probe compatible with many high-quality temperature readouts and resistance measurement systems.
Four-wire technology reduces the influence of the lead wires.
The Inconel sheath supports reliable use up to 670 °C.
With two lengths available, the sensor can also be adapted to different calibration baths, furnaces and fixed-point cells.
Typical Calibration Applications with the Additel AccuMac AM 1760
The Additel AccuMac AM 1760 is particularly suitable as a reference probe for:
- Pt100 calibrations
- PRT calibrations
- Reference thermometer calibrations
- Comparison calibrations
- Fixed-point measurements
- Triple point of water measurements
- Calibration baths
- Liquid baths
- Temperature furnaces
- Dry-block calibrators with suitable uniformity
- Stability measurements
- Temperature comparison measurements
- Research and development
In a comparison calibration, the AM 1760 is placed together with the device under test in a stable and homogeneous temperature source.
The resistance of the reference probe is measured with a suitable high-precision reference readout. The individual calibration coefficients then allow the reference temperature to be determined.
At europascal you can obtain the Additel AccuMac AM 1760 as well as suitable temperature readouts, calibration baths, dry-block calibrators and accessories for professional temperature calibration.
You can find further suitable products among the temperature sensors from europascal. For traceable calibration of your temperature probes and reference thermometers, the europascal Calibration Service is at your disposal.





