europascal BH2-3000B Precise High Pressure Deadweight Tester

The europascal BH2-3000B High Pressure Deadweight Tester is a precision single-piston hydraulic pressure standard for professional calibration applications from 40 to 3,000 bar (300 MPa). With a standard accuracy of 0.02 % of reading and enhanced performance up to 0.008 % on request, the BH2-3000B features a tungsten carbide piston and cylinder for reliable high-pressure operation. Optional motorization, piston position visualization and automatic pressure calculation provide additional flexibility for demanding calibration laboratories.

The europascal BH2-3000B High Pressure Deadweight Tester is a high-performance hydraulic pressure standard designed for precise calibration applications from 40 to 3,000 bar. With a standard accuracy of 0.02 % of reading and enhanced performance up to 0.008 % on request, the BH2-3000B provides reliable physical pressure generation for professional calibration laboratories and demanding industrial pressure measurement applications.

The instrument covers a measuring range from 40 to 3,000 bar, corresponding to 4 to 300 MPa, 4,000 to 300,000 kPa and 600 to 44,000 psi according to the manufacturer.

The BH2-3000B uses a single precision piston-cylinder system. Both the piston and cylinder are manufactured from tungsten carbide, providing excellent hardness, wear resistance and dimensional stability for high-pressure operation.

The instrument operates hydraulically with oil. The reference pressure is physically generated by calibrated stainless steel masses acting on the precision piston-cylinder assembly.

The robust BH2 platform is designed for regular or intensive calibration work and can additionally be equipped with optional motorization, piston position visualization and automatic pressure calculation.

The europascal calibration service provides professional calibration solutions for deadweight testers, pressure balances, pressure gauges, pressure sensors, transmitters and other pressure measuring equipment.

Your Benefits at a Glance

The europascal BH2-3000B High Pressure Deadweight Tester provides:

  • Hydraulic high-pressure deadweight tester
  • Measuring range from 40 to 3,000 bar
  • Measuring range from 4 to 300 MPa
  • Measuring range from 4,000 to 300,000 kPa
  • Manufacturer range from 600 to 44,000 psi
  • Standard accuracy of 0.02 % of reading
  • Enhanced accuracy up to 0.008 % on request
  • Single-piston configuration
  • One precision piston-cylinder system
  • Tungsten carbide piston
  • Tungsten carbide cylinder
  • Hydraulic operation with oil
  • Stainless steel calibration masses
  • Manual rotation of the masses as standard
  • Standard or customer-specific gravity
  • Optional motorized mass rotation
  • Optional piston position visualization
  • Optional automatic pressure calculation
  • Optional computer data transfer
  • Robust construction for professional calibration laboratories

Measuring Range from 40 to 3,000 bar

The europascal BH2-3000B covers a hydraulic measuring range from:

40 to 3,000 bar

This corresponds to:

4 to 300 MPa

and:

4,000 to 300,000 kPa

AREMECA specifies the imperial measuring range as:

600 to 44,000 psi

The maximum pressure is therefore:

3,000 bar / 300 MPa

This wide pressure range makes the BH2-3000B particularly suitable for laboratories that need to calibrate both medium-high and very high-pressure measuring instruments with a single pressure balance.

Accuracy of 0.02 % of Reading

The standard accuracy of the europascal BH2-3000B is:

0.02 % of reading

For applications with more demanding metrological requirements, AREMECA specifies enhanced performance:

up to 0.008 % on request

The specification as a percentage of reading is particularly advantageous for an instrument with such a wide measuring range.

The actual measurement uncertainty achieved during calibration additionally depends on the individual calibration certificate, operating conditions and complete uncertainty budget.

Physical Piston Gauge Measurement Principle

The europascal BH2-3000B High Pressure Deadweight Tester operates according to the physical piston gauge principle.

The fundamental relationship is:

p = F / A

where:

p = generated pressure
F = force acting on the piston
A = effective piston area

The gravitational force generated by the calibrated masses is primarily determined by:

F = m × g

where:

m = applied mass
g = gravitational acceleration

The simplified pressure relationship is therefore:

p = (m × g) / A

This direct physical relationship between mass, gravity and effective piston area is the basis of deadweight tester pressure generation.

For precision calibration, additional correction quantities are considered according to the calibration certificate and operating conditions.

Single-Piston Configuration

The europascal BH2-3000B is designed as a:

single-piston deadweight tester

The instrument incorporates:

1 precision piston-cylinder system

The complete specified measuring range from 40 to 3,000 bar is covered with this single piston configuration.

This provides straightforward operation because there is no need to switch between separate low-pressure and high-pressure piston-cylinder assemblies within the specified range.

Tungsten Carbide Piston and Cylinder

The BH2-3000B uses:

Piston: tungsten carbide

Cylinder: tungsten carbide

Tungsten carbide provides high hardness, excellent wear resistance and good dimensional stability.

These characteristics are particularly important for a precision piston-cylinder assembly operating at pressures up to:

3,000 bar / 300 MPa

The effective piston area is one of the most important metrological parameters of a deadweight tester because it directly determines the generated pressure.

Hydraulic Operation with Oil

The europascal BH2-3000B is a hydraulic pressure balance operating with:

oil

Hydraulic operation is particularly suitable for generating very high pressures because liquids have relatively low compressibility compared with gases.

Correct filling and venting are important for stable pressure generation.

Trapped gas can increase settling times and affect pressure stability. Cleanliness is also important because contamination can adversely affect precision hydraulic components.

Compatibility between the hydraulic medium and the connected device under test should always be verified before calibration.

High-Pressure Calibration up to 300 MPa

A maximum pressure of 3,000 bar corresponds to:

300 MPa

At these pressures, every pressure-containing component of the calibration setup must be suitable for the maximum applied pressure.

This includes:

  • High-pressure tubing
  • Fittings
  • Adapters
  • Valves
  • Seals
  • Pressure generator
  • Device under test
  • Associated safety components

The system must be completely depressurized before hydraulic connections are changed.

Measurement Uncertainty Contributions

The complete measurement uncertainty of a calibration does not depend only on the stated accuracy of the pressure balance.

Relevant influence quantities can include:

  • Calibration uncertainty of the masses
  • Effective piston area
  • Pressure coefficient
  • Elastic deformation of the piston-cylinder assembly
  • Local gravitational acceleration
  • Air buoyancy
  • Piston-cylinder temperature
  • Thermal expansion
  • Hydraulic fluid properties
  • Hydrostatic head difference
  • Repeatability
  • Sensitivity
  • Pressure stability
  • Resolution of the device under test
  • Environmental conditions

At pressures up to 300 MPa, pressure-related changes in the effective piston area can become particularly important.

The final measurement uncertainty should therefore be determined from the actual calibration data and calibration procedure.

Stainless Steel Calibration Masses

The BH2 series uses:

stainless steel calibration masses

The calibrated masses generate the gravitational force acting on the piston.

Together with the effective piston area and gravitational acceleration, they determine the generated reference pressure.

The mass sets are ergonomically designed for loading onto the pressure balance.

Different pressure units can be selected according to the required configuration.

Manual Mass Rotation as Standard

The BH2 platform uses manual rotation of the masses as standard.

Rotation of the piston helps minimize friction effects between the piston and cylinder and supports stable operation of the precision piston-cylinder assembly.

For laboratories performing frequent or repetitive calibration sequences, motorized rotation is available as an option.

Standard or Local Gravity

The force generated by calibration masses depends directly on gravitational acceleration.

The conventional standard value is:

g = 9.80665 m/s²

Actual gravitational acceleration varies geographically.

The BH2-3000B can therefore be configured for:

standard gravity

or:

customer-specific local gravity

Using the correct gravity value is important for precision pressure calibration because a relative change in gravitational acceleration directly changes the force produced by the calibration masses.

Environmental Conditions

AREMECA specifies an operating temperature range of:

18 to 28 °C

The manufacturer specifies calibration reference conditions including:

Temperature: 20 °C

Atmospheric pressure: 1013.25 hPa

Relative humidity: 50 %

Gravitational acceleration: 9.80665 m/s²

If the instrument is used under different environmental conditions, the generated pressure should be calculated using the relevant correction quantities.

Leveling and Installation

Correct leveling is important for reliable piston gauge operation.

The BH2 platform therefore incorporates:

  • Leveling bubble
  • Adjustable feet

The pressure balance should be installed on a stable work surface and carefully leveled before calibration.

Correct leveling helps the piston move vertically and minimizes unwanted lateral forces within the piston-cylinder assembly.

Device Under Test Connection

The BH2 series provides a connection for the instrument under test using:

swivel G1/2 HP

Other fittings are available as options.

The appropriate adapter must always be selected according to the pressure rating, thread type and sealing system of the device under test.

At pressures up to 3,000 bar, the pressure rating of every fitting and adapter must be verified before operation.

Instrument Weight

The standard BH2-3000B weighs approximately:

22 kg without calibration masses

When equipped with option OP0004 or OP0301, AREMECA specifies approximately:

24 kg without calibration masses

The total system weight additionally depends on the selected calibration mass set.

Optional OP0004 Motorization

The europascal BH2-3000B can be equipped with:

Option OP0004

This provides:

  • Motorized rotation of the masses
  • Piston position visualization using indicator lights

Motorized rotation reduces the need for manual intervention during repetitive calibration procedures.

Piston position visualization assists the operator in maintaining the piston within the correct working position.

Optional OP0301 Calculator Module

For more advanced calibration workflows, the BH2-3000B can be equipped with:

Option OP0301

This option provides additional functions such as:

  • Motorized operation
  • Dynamic piston position visualization
  • Environmental sensors
  • Automatic pressure calculation
  • Automatic pressure display
  • Consideration of environmental quantities
  • Calibration data traceability
  • Computer data transfer

This allows the standard manually operated BH2 platform to be expanded toward a more automated calibration workflow.

Technical Highlights

Model: europascal BH2-3000B

Device type: High Pressure Hydraulic Deadweight Tester

Configuration: Single piston

Number of pistons: 1

Measuring range: 40 to 3,000 bar

Measuring range: 4 to 300 MPa

Measuring range: 4,000 to 300,000 kPa

Manufacturer imperial range: 600 to 44,000 psi

Maximum pressure: 3,000 bar / 300 MPa

Standard accuracy: 0.02 % of reading

Enhanced accuracy: Up to 0.008 % on request

Piston material: Tungsten carbide

Cylinder material: Tungsten carbide

Pressure medium: Oil

Instrument weight without masses: Approx. 22 kg

Weight with OP0004/OP0301 without masses: Approx. 24 kg

Connection: Swivel G1/2 HP

Measurement principle: Piston gauge / deadweight tester

Operating temperature: 18 to 28 °C

Standard gravity: 9.80665 m/s²

Local gravity: Available

Motorization: Optional OP0004 / OP0301

Automatic pressure calculation: Optional OP0301

Computer data transfer: Optional OP0301

Technical Data

Feature europascal BH2-3000B
Manufacturer europascal Aremeca
Model BH2-3000B
Device type High Pressure Hydraulic Deadweight Tester
Configuration Single piston
Number of pistons 1
Measuring range 40–3,000 bar
Measuring range 4–300 MPa
Measuring range 4,000–300,000 kPa
Manufacturer imperial range 600–44,000 psi
Maximum pressure 3,000 bar / 300 MPa
Standard accuracy 0.02 % of reading
Enhanced accuracy Up to 0.008 % on request
Piston material Tungsten carbide
Cylinder material Tungsten carbide
Pressure medium Oil
Weight without masses Approx. 22 kg
DUT connection Swivel G1/2 HP
Operating temperature 18–28 °C
Standard gravity 9.80665 m/s²
Local gravity Available
Mass rotation Manual as standard
Motorization Optional
Piston position visualization Optional
Automatic pressure calculation Optional OP0301
Computer data transfer Optional OP0301

Position within the BH2 High-Pressure Series

The BH2 series provides several single-piston configurations for high hydraulic pressures.

Model Pressure Range
BH2-2000B 40–2,000 bar
BH2-2500B 40–2,500 bar
BH2-3000B 40–3,000 bar
BH2-4000B 40–4,000 bar
BH2-5000B 40–5,000 bar
BH2-6000B 200–6,000 bar
BH2-7000B 200–7,000 bar

The BH2-3000B therefore provides a particularly useful combination of a relatively low starting pressure of 40 bar and a maximum pressure of 3,000 bar.

BH2-3000B vs. BH2-4000B

The BH2-3000B and BH2-4000B address similar high-pressure calibration applications but differ in maximum pressure.

Feature BH2-3000B BH2-4000B
Minimum pressure 40 bar 40 bar
Maximum pressure 3,000 bar 4,000 bar
Maximum pressure 300 MPa 400 MPa
Standard accuracy 0.02 % 0.02 %
Number of pistons 1 1
Piston material Tungsten carbide Tungsten carbide
Cylinder material Tungsten carbide Tungsten carbide
Pressure medium Oil Oil
Standard operation Manual Manual
Motorization Optional Optional
Automatic calculation Optional Optional

The BH2-3000B is therefore an excellent choice when the required calibration range does not exceed 300 MPa, while the BH2-4000B extends the maximum pressure to 400 MPa.

Typical Calibration Applications

The europascal BH2-3000B High Pressure Deadweight Tester is suitable for testing, adjustment and calibration of:

  • High-pressure gauges
  • Digital pressure gauges
  • Pressure sensors
  • Pressure transducers
  • Pressure transmitters
  • Pressure switches
  • Reference pressure instruments
  • Hydraulic high-pressure measurement systems
  • Industrial pressure measuring equipment

Typical users include:

  • Professional calibration laboratories
  • Pressure instrument manufacturers
  • Industrial quality assurance departments
  • Research laboratories
  • High-pressure technology facilities
  • Calibration service organizations

Calibration and Traceability

Regular calibration of a high-pressure deadweight tester is important for maintaining reliable metrological performance.

Important parameters include:

  • Effective piston area
  • Pressure coefficient
  • Piston-cylinder characteristics
  • Calibration mass values
  • Local gravitational acceleration
  • Temperature
  • Repeatability
  • Relevant correction quantities

The europascal calibration service provides professional calibration solutions for pressure balances, deadweight testers, high-pressure gauges, pressure sensors and pressure transmitters.

europascal can also provide support with maintenance, configuration and selection of suitable high-pressure calibration equipment.

Why Choose the europascal BH2-3000B?

The europascal BH2-3000B Precise High Pressure Deadweight Tester combines a broad hydraulic pressure range from 40 to 3,000 bar with a standard accuracy of 0.02 % of reading and enhanced performance up to 0.008 % on request.

Its single-piston configuration and tungsten carbide piston-cylinder assembly provide a robust physical pressure reference for demanding calibration applications up to 300 MPa.

Manual operation provides a straightforward and reliable calibration platform, while optional motorization, piston position visualization, automatic pressure calculation and computer data transfer allow the BH2-3000B to be adapted to more advanced laboratory workflows.

For professional laboratories requiring a physical hydraulic reference covering pressures from tens of bar to 3,000 bar, the BH2-3000B provides an excellent combination of measuring range, metrological performance, robust construction and configurable automation.

Looking for the europascal BH2-3000B or a hydraulic high-pressure calibration solution up to 3,000 bar? Contact europascal for expert advice, system configuration and professional calibration services tailored to your requirements.

Further products from europascal Aremeca can be found here.

Manufacturer information about the europascal BH2-3000B can be found on the official Aremeca Instrumentation website.

Download Datasheet

Medium

Druckbereich

Koben-zylindereinheit

Genauigkeit

Motorisierung

Other Products

Scroll to Top

Request a Quote Now!

Download Product Datasheet

Enter your email address below to access the official product datasheet, including technical specifications, measurement ranges, product features, and performance details.