europascal

Fluke Calibration 2271A Modular Pressure Calibrator

Fluke Calibration 2271A is a modular automated pressure calibrator for professional calibration of pressure gauges, sensors, switches and transmitters from -100 kPa to 20 MPa (-15 to 3,000 psi). Interchangeable PM200 and PM500 modules provide performance down to 0.02% FS or 0.01% of reading, while the integrated Electrical Measurement Module offers mA and DC voltage measurement, 24 V loop power and HART communication. Automated pressure control, dual test ports and Ethernet, RS232 and USB interfaces make the 2271A ideal for efficient laboratory and industrial calibration.

Fluke Calibration 2271A is a modular automated pressure calibrator designed for professional calibration of pressure gauges, pressure transmitters, pressure switches and pressure sensors. With a pressure measurement range from -100 kPa to 20 MPa (-15 psi to 3,000 psi), interchangeable PM200 and PM500 pressure measurement modules, integrated electrical measurement functions and automated pressure control, the 2271A provides a complete solution for industrial and laboratory pressure calibration. Fluke

The modular architecture is one of the key strengths of the Fluke Calibration 2271A. Two pressure measurement modules can be installed in the chassis, allowing different pressure ranges and accuracy classes to be combined in one system. As calibration requirements change, modules can be exchanged or additional ranges purchased without replacing the complete controller. PM200 modules are designed for economical industrial calibration with specifications down to 0.02% FS, while PM500 modules provide measurement uncertainty down to 0.01% of reading over significant portions of their ranges. Fluke

The 2271A also integrates an Electrical Measurement Module, or EMM, with mA and DC voltage measurement, 24 V DC loop power and HART communication. This allows pressure transmitters to be connected, powered, controlled and calibrated directly with the same system. Fully automated closed-loop calibration of 4–20 mA smart pressure transmitters can therefore be performed without requiring a separate digital multimeter, loop supply or external HART resistor. Fluke

For calibration laboratories looking to improve throughput, the 2271A can automate pressure generation, stabilization, electrical measurement and evaluation. It can also be controlled by Fluke Calibration COMPASS for Pressure software or integrated into custom automated calibration systems through Ethernet, RS232 and USB interfaces. Fluke

europascal supports professional users in selecting suitable pressure measurement modules, pressure connections and accessories for the 2271A. europascal also operates one of its core accredited calibration capabilities in pressure metrology, covering a very wide range of pressure instruments. For traceable calibration and service, visit the europascal Calibration Service. europascal is additionally listed by Fluke as a regional service provider for Fluke Calibration pressure and flow products. europascal.de

Your Benefits at a Glance

  • Automated pneumatic pressure calibrator
  • Pressure range from -100 kPa to 20 MPa
  • Equivalent range from -15 psi to 3,000 psi
  • Modular pressure measurement concept
  • Two pressure measurement modules can be installed
  • PM200 pressure measurement modules
  • PM500 pressure measurement modules
  • Accuracy classes down to 0.02% FS or 0.01% of reading
  • Modules can be exchanged as calibration requirements change
  • Dynamic control precision down to 0.001% of range span
  • Typical control turndown of 10:1
  • Low control point down to 1 kPa absolute
  • Automatic pressure generation and control
  • Integrated Electrical Measurement Module
  • DC current measurement
  • DC voltage measurement
  • 24 V loop power
  • Integrated HART communication
  • Built-in 250 Ω HART resistor
  • Fully automated pressure transmitter calibration
  • Dual test ports for multiple devices under test
  • Optional Contamination Prevention System
  • Ethernet interface
  • RS232 interface
  • USB interface
  • Compatible with COMPASS for Pressure
  • Suitable for automated calibration laboratories
  • Suitable for industrial pressure calibration
  • Professional calibration and service through europascal

Fluke positions the 2271A as a complete automated system for calibrating pressure gauges and sensors, with modular measurement capabilities and electrical transmitter calibration functions in one chassis. Fluke

Ideal for Demanding Calibration Applications

The Fluke Calibration 2271A is designed for laboratories and industrial facilities that need to calibrate a broad variety of pressure instruments with repeatable automated processes.

Typical applications include:

  • Calibration of analog pressure gauges
  • Calibration of digital pressure gauges
  • Pressure transmitter calibration
  • HART pressure transmitter calibration
  • 4–20 mA pressure transmitter testing
  • Pressure switch testing
  • Pressure sensor calibration
  • Production testing
  • Quality assurance
  • Automated calibration stations
  • Calibration laboratory applications
  • Industrial maintenance
  • Research and development

The system is particularly attractive for laboratories that are expanding their pressure calibration capabilities because its modular design allows the available workload to grow without replacing the complete pressure controller. Fluke

Pressure Range up to 20 MPa

The Fluke Calibration 2271A covers pressure measurement ranges from:

-100 kPa

to:

20 MPa

or approximately:

-1 bar to 200 bar.

In imperial units, Fluke specifies:

-15 psi to 3,000 psi. Fluke

The broad range makes the 2271A suitable for both relatively low pneumatic pressures and demanding industrial pressure calibration applications.

The overall working range depends on the pressure measurement modules installed in the system. This means the instrument can be configured specifically for the customer’s actual calibration workload instead of relying on one fixed sensor range.

PM200 Pressure Measurement Modules

The PM200 family provides an economical solution for many industrial pressure calibration tasks.

For most commonly used PM200 ranges, Fluke specifies:

0.02% FS.

This level of performance is particularly suitable for:

  • Analog pressure gauges
  • Lower-accuracy pressure transmitters
  • Pressure switches
  • General industrial pressure instruments

The PM200 modules use rugged silicon pressure sensing technology and are designed to provide fast and reliable pressure control. Fluke

Multiple PM200 ranges are available, extending from very low bidirectional pressures through gauge pressure ranges up to 20 MPa / 3,000 psi. Individual accuracy specifications depend on the selected module, particularly for the lowest ranges. Fluke

PM500 Pressure Measurement Modules

For more demanding calibration requirements, the Fluke Calibration 2271A also supports PM500 pressure measurement modules.

PM500 modules use highly characterized and linearized silicon pressure sensors.

For many ranges, Fluke specifies a measurement uncertainty of:

0.01% of reading

between approximately 50% and 100% of the range. Fluke

This makes PM500 modules suitable for higher-accuracy pressure transmitters, digital pressure gauges and other devices requiring better measurement performance than the standard PM200 class.

By combining PM200 and PM500 modules, a laboratory can match both accuracy and investment cost to its actual workload.

Modular Design

The modular design is one of the most important characteristics of the Fluke Calibration 2271A.

Two measurement modules can be installed in the same chassis. Modules can be selected according to pressure range and required performance, then replaced when calibration requirements change. Fluke

This provides several advantages.

A laboratory can start with only the pressure ranges it currently requires.

Additional ranges can be added later.

A module can be replaced without purchasing a new controller.

Different accuracy classes can be selected for different workloads.

The investment therefore remains adaptable as the number and type of devices under test change.

Dynamic Pressure Control

The Fluke Calibration 2271A not only measures pressure but also automatically generates and regulates it when supplied with an appropriate pneumatic pressure source.

For most ranges, Fluke specifies dynamic control precision of:

0.001% of range span.

Special low-pressure configurations have different values. The PM200-BG2.5K is specified at 0.005% of range span, while certain very low-pressure PM500 configurations are specified at 0.002% of range span. Fluke

This precise regulation enables the 2271A to establish repeatable calibration points automatically.

Control Turndown of 10:1

The typical control turndown is:

10:1. Fluke

Fluke notes that the available supply pressure is an important factor in achieving the specified control performance. To meet the control specification, the supply pressure should not be greater than approximately ten times the range of the active measurement module.

This is particularly relevant when very different pressure modules are installed in the same controller.

Fast Pressure Settling

The settling time depends on the installed module and the calibration setup.

Fluke specifies typical settling times including approximately:

20 seconds for many PM200 modules

and approximately:

30 to 45 seconds for PM500 modules depending on range. Fluke

Actual settling time can also be influenced by test volume, pressure tubing, leaks, supply pressure and vacuum performance.

Fast and repeatable pressure settling is important in automated calibration because it directly affects total calibration time.

Integrated Electrical Measurement Module

The Fluke Calibration 2271A integrates pressure and electrical calibration functions.

Its Electrical Measurement Module enables measurements required for pressure transmitter calibration without the need for several separate instruments.

The module supports electrical measurements through standard 4 mm connections and allows signals up to 30 V DC relative to chassis ground. europascal.de

The integrated concept simplifies calibration bench design and reduces setup time when pressure transmitters are regularly calibrated.

HART Pressure Transmitter Calibration

The EMM provides integrated HART communication.

It also includes an internal:

250 Ω resistor

that can be switched into the circuit when required for HART communication. Fluke

This eliminates the need for an external resistor in many HART calibration setups.

The 2271A can therefore support fully automated closed-loop calibration of smart pressure transmitters.

Pressure is generated and controlled automatically, the transmitter is supplied with loop power, its electrical signal is measured and the calibration procedure can be evaluated automatically.

24 V Loop Power

The integrated Electrical Measurement Module also provides:

24 V DC loop power.

This allows compatible two-wire pressure transmitters to be powered directly from the Fluke Calibration 2271A during calibration. Fluke

A typical transmitter calibration setup can therefore use the 2271A for:

pressure generation,

reference pressure measurement,

24 V transmitter supply,

mA measurement,

and HART communication.

This significantly reduces external instrumentation and cabling.

Dual Test Ports

The Fluke Calibration 2271A incorporates dual test ports, allowing multiple devices under test to be connected conveniently. Fluke

This can increase laboratory throughput when multiple pressure instruments need to be connected to the same pressure source.

The test-port concept is also designed to support convenient connection and removal of devices without unnecessarily complicated adapter arrangements.

Different chassis configurations are available with NPT, BSP and alternative test-port arrangements. Fluke

Pneumatic Pressure Calibration

The Fluke Calibration 2271A is a pneumatic pressure calibrator.

Fluke specifies the pressure supply medium as:

clean, dry air

or:

nitrogen.

For nitrogen, industrial grade with approximately 99.5% purity is specified. Fluke

Fluke also places limits on particulate contamination, moisture and hydrocarbon content to protect the internal pneumatic system.

A clean pneumatic supply is especially important for maintaining controller reliability and protecting the pressure measurement modules.

Optional Contamination Prevention System

Pressure instruments returning from industrial service may contain oil, moisture, particles or other contamination.

To protect the controller from contamination entering through the device under test, Fluke offers an optional Contamination Prevention System, or CPS.

The CPS uses a combination of controlled flow direction, a gravity sump and filtration to reduce the risk of contamination reaching the pressure controller. Fluke

This option can be particularly valuable in calibration laboratories handling pressure gauges and transmitters from different customer applications.

Automated Pressure Transmitter Calibration

Automation is one of the strongest capabilities of the Fluke Calibration 2271A.

Fluke describes workflows in which the calibrator automatically moves through predefined pressure points, waits for the selected dwell time, measures the transmitter output and evaluates the result against a defined tolerance. Fluke

The user can define:

  • Start pressure
  • End pressure
  • Number of test points
  • Direction of calibration
  • Dwell time
  • Maximum settling time
  • DUT zero and full-scale values
  • Pass/fail tolerance

Once started, the procedure can run in closed-loop mode with minimal operator intervention.

This helps improve consistency between calibrations while allowing laboratory personnel to perform other tasks.

COMPASS for Pressure

The Fluke Calibration 2271A can be integrated with COMPASS for Pressure software for automated calibration and test sequences.

COMPASS can support automated calibration of single or multiple devices under test and is designed specifically for pressure calibration applications. Fluke

The instrument also offers a full remote interface for integration into customer-specific software environments.

Ethernet, RS232 and USB

For automated operation, the 2271A provides:

Ethernet

RS232

and:

USB. europascal.de

These interfaces provide flexibility when integrating the controller into an existing calibration laboratory, production test system or custom software environment.

Technical Highlights

Model: Fluke Calibration 2271A

Device type: Modular automated pressure calibrator

Pressure range: -100 kPa to 20 MPa

Imperial range: -15 psi to 3,000 psi

Measurement modules: PM200 and PM500

Installed modules: up to two pressure measurement modules

PM200 performance: down to 0.02% FS for most industrial ranges

PM500 performance: down to 0.01% of reading for many ranges

Dynamic control precision: down to 0.001% of range span

Control turndown: 10:1 typical

Low control point: 1 kPa absolute

Electrical module: integrated EMM

Loop supply: 24 V DC

Communication with smart transmitters: HART

Remote interfaces: Ethernet, RS232, USB

Test ports: dual test ports

Pressure medium: clean dry air or nitrogen

Optional contamination protection: CPS

Warm-up time: approximately 15 minutes

Chassis weight: approximately 15 kg

The specifications correspond to current Fluke manufacturer information and the 2271A documentation hosted by europascal. Fluke

Technical Data

Feature Fluke Calibration 2271A
Manufacturer Fluke Calibration
Model 2271A
Device type Automated Pressure Calibrator
Pressure range -100 kPa to 20 MPa
Pressure range in psi -15 to 3,000 psi
Maximum range in bar Approx. 200 bar
Measurement modules PM200, PM500
Module positions 2
PM200 performance Down to 0.02% FS
PM500 performance Down to 0.01% reading
Dynamic control precision Down to 0.001% range span
Control turndown 10:1 typical
Low control point 1 kPa absolute
Electrical measurement Integrated EMM
Loop supply 24 V DC
HART Yes
Test ports Dual
Remote interfaces Ethernet, RS232, USB
Supply medium Clean dry air or nitrogen
Maximum supply pressure 23 MPa gauge
Maximum test-port pressure 20 MPa absolute
Operating temperature 15 °C to 35 °C
Warm-up Approx. 15 min
Chassis weight Approx. 15 kg
Width 442 mm
Depth 446 mm
Protection IP20

The exact dimensions depend on the selected HC20 or P3000 test-port chassis configuration. europascal.de

Why Choose the Fluke Calibration 2271A?

The Fluke Calibration 2271A combines automatic pneumatic pressure control, modular reference measurement and electrical transmitter calibration in one professional platform.

Its pressure range from -100 kPa to 20 MPa covers a large proportion of industrial pneumatic calibration workloads. Two interchangeable module positions allow users to combine different ranges, while the choice between PM200 and PM500 modules provides flexibility between economical industrial calibration and higher-accuracy applications.

The integrated EMM makes the system particularly attractive for calibration laboratories working with pressure transmitters. mA and voltage measurement, 24 V loop power and integrated HART communication mean that many transmitter calibrations can be completed without additional electrical instrumentation.

Automated pressure control, programmable procedures and compatibility with COMPASS for Pressure can reduce repetitive manual work and improve consistency in calibration processes.

The optional CPS also helps protect the controller when pressure instruments returning from industrial service may contain contamination.

For laboratories seeking a modular pressure calibration platform that can grow with changing workloads, the Fluke Calibration 2271A provides a flexible and scalable solution.

europascal offers professional support for Fluke Calibration pressure equipment and is listed by Fluke as a regional service provider for pressure and flow calibration products. Fluke

Typical Applications with the Fluke Calibration 2271A

The Fluke Calibration 2271A is particularly suitable for automated calibration of pressure gauges, pressure transmitters, digital pressure indicators, pressure switches and pressure sensors.

Its integrated HART functionality makes it especially useful for modern smart pressure transmitters, while the modular PM200 and PM500 concept enables calibration laboratories to match pressure range and accuracy to their workload.

For complete pressure calibration workstations, europascal can support the selection of suitable measurement modules, pressure supply components, fittings and accessories.

For traceable calibration of your pressure equipment, visit the europascal Calibration Service.

Further professional pressure calibration equipment can be found in the pressure technology section at europascal. europascal currently lists a broad range of pressure controllers, calibrators, pressure modules and reference instruments. europascal.de

Further products from Fluke Calibration can be found here. europascal lists Fluke Calibration as one of its core manufacturers for professional calibration technology and identifies itself as an authorized service center for Fluke Calibration pressure and flow products in the EMEA region. europascal.de

Manufacturer information about the Fluke Calibration 2271A can be found on the official Fluke Calibration website.

Download Datasheet

Medium

Druckbereich

Genauigkeit

Datenlogger

Schnittstelle

Strom-versorgung

Manometer Typen

HART-Kommunikation

Bauart

Kanäle

Other Products

Scroll to Top

Request a Quote Now!

Download Product Datasheet

Enter your email address below to access the official product datasheet, including technical specifications, measurement ranges, product features, and performance details.